¢Ã Test voltages of 250V, 500 V, 1000 V, 2500 V, and 5000 V ¢Ã Capable of testing in 50V increments between 250V and 1000V and 100V increments between 1000V and 5000V ¢Ã Measures 0 to 1 Tera-Ohm ¢Ã Warning voltage function alerts the user that line voltage is present and gives the voltage reading up to 600 V AC or DC ¢Ã Guard system eliminates the effect of surface leakage current on high-resistance measurements ¢Ã Large digital/analog LCD shows detailed measurement data ¢Ã Cable or insulation capacitance ¢Ã Leakage current ¢Ã Ramp function (0-5000 V DC) for breakdown testing ¢Ã Timer 1 to 99 minutes ¢Ã Polarization index and dielectric absorption calculated automatically ¢Ã 99 memory locations store all measurements parameters ¢Ã Includes Quicklink 1550B Software and Optical Interface cable ¢Ã Two year warranty
Specifications
Test Voltage 250 V DC
Range:
200 k¥Ø to 5 G¥Ø
Accuracy:
5%
Range:
5 G¥Ø to 50 G¥Ø
Accuracy:
20%
Test Voltage 500 V DC
Range:
200 k¥Ø to 10 G¥Ø
Accuracy:
5%
Range:
10 G¥Ø to 100 G¥Ø
Accuracy:
20%
Test Voltage 1000 V DC
Range:
200 k¥Ø to 20 G¥Ø
Accuracy:
5%
Range:
20 G¥Ø to 200 G¥Ø
Accuracy:
20
Test Voltage 2500 V DC
Range:
200 k¥Ø to 50 G¥Ø
Accuracy:
5%
Range:
50 G¥Ø to 500 G¥Ø
Accuracy:
20%
Test Voltage 5000 V DC
Range:
200 k¥Ø to 100 G¥Ø
Accuracy:
5%
Range:
100 G¥Ø to 1 T¥Ø
Accuracy:
20%
Short circuit current
> 1 mA and < 2 mA
Bar graph
Range:
0 to 1 T¥Ø
Insulation test voltage
Accuracy:
0% to 10% at 1 mA load current
Induced AC mains current rejection
Max:
2 mA
Charging rate for capacitive load
5 s/µF
Leakage current
Range:
1 nA to 2 mA
Accuracy:
¡¾(5% + 2 nA)
Capacitance measurements
Range:
0.01 µF to 15.00 µF
Accuracy:
¡¾(15% rdg + 0.03 µF)
Live circuit indicator
Range:
30 V to 600 V (AC/DC, 50/60 Hz)
Accuracy:
¡¾(5% + 2V)
Timer
1 - 99 minutes
Settable:
1 minute increments
Indication:
1 second resolution
Ramp
0% to 100% of selected test voltage, or until breakdown
Environmental Specifications
Operating Temperature
-20¡ÆC to +50¡ÆC (-4¡ÆF to 122¡ÆF)
Storage Temperature
-20 ¡ÆC to +65 ¡ÆC (-4¡ÆF to 149¡ÆF)
Humidity (Without Condensation)
80% at 31 ¡ÆC (87.8¡ÆF), 50% at 50 ¡ÆC (122¡ÆF)
Dust/water resistance
IP-40
Operating Altitude
2,000 m
Safety Specifications
Safety Rating
EN 61010-1, CAT III 600V, EN61557, Parts 1 & 2
EMI immunity
EN 61326
Mechanical & General Specifications
Size
170mm x 242mm x 330 mm (6.7" x 9.5" x 13.0")
Weight
4 kg (with battery) (8.8 lbs)
Batteries
12 V, lead-acid, rechargeable
Warranty
2 years
Battery Life
Test voltage:
Number of tests
250 V:
4100
500 V:
3600
1000 V:
3200
2500 V:
2500
5000 V:
1500
Shock & Vibration
MIL-PRF-28800F, Class II
Charger input
AC:
85 to 250 V AC, 50/60 Hz, 50 VA
Included Accessories
- Test leads and low-leakage probes -Alligator Clips - Interface adapter and cable - Quicklink 1550B PC software - Line cord - Soft carrying case with waterproof bottom - User's manual